This chamber has uniformly heated glass bottom with large window of 42 mm diameter. Can be used with tissue slices and samples attached to a cover slip. Separate compartments for inflow and outflow prevent bubbles from entering the chamber and provide smooth perfusion. Can be used with controlled flow perfusion systems. Might need an objective heater if water immersion optics is used. Requires a temperature controller. Mounted on a 128x86 mm frame, 5 mm thick, which fits most microscope stages. Might require a microscope adapter
Specify microscope model when ordering
|
Specifications:
|
|
Dimensions:
|
42 mm diameter optical aperture of conical shape to provide effective width of 50 mm, 3 mm thick
|
|
Low-profile:
|
1 mm heated glass, mounted on 5 mm thick 128x86 mm frame
|
|
Temperature stability:
|
0.01°C, built-in sensor
|
|
Glass bottom:
|
1 mm thick
|